Photon energy
1 eV = 8065.5 cm⁻¹ = 241.8 THz = 96.49 kJ mol⁻¹. Thermal energy at 298 K: kBT = 207 cm⁻¹ = 2.48 kJ mol⁻¹ = 25.7 meV. Transitions far above kBT start from the ground state; those below it are thermally populated.
A field atlas for doctoral students in biophysics
Thirty-two techniques arranged by the physical process they exploit, each with its governing equations, what it reveals about biomolecules, and where it stops working. Select any technique, on the map or on the spectrum, to open its full card. Four live calculators and a question-to-method guide close the page.
The hierarchy · levels 1 to 3 shown here, level 4 inside each card
The first split is physical: what happens to the photon (or particle) at the sample. The second split is what transition or observable is read out. Cross-links in each card connect techniques that share a transition, such as IR and Raman for vibrations.
Energy axis
Bars mark the energy of the transition being probed, not necessarily the photon used. Raman is plotted at vibrational energies although it uses visible lasers; neutron spectroscopy is plotted at its energy transfers. Mass, surface and force methods without a spectral axis are omitted.
Logarithmic wavelength axis, 1 pm to 1 km. Region boundaries are conventional and approximate. Energy conversion: E[eV] = 1239.84 / λ[nm]. Select a bar to open the technique.
Shared physics
Each technique card assumes these. They explain why selection rules exist, why NMR is insensitive, why fluorescence is fast and phosphorescence slow, and why so many methods end up measuring a correlation function.
1 eV = 8065.5 cm⁻¹ = 241.8 THz = 96.49 kJ mol⁻¹. Thermal energy at 298 K: kBT = 207 cm⁻¹ = 2.48 kJ mol⁻¹ = 25.7 meV. Transitions far above kBT start from the ground state; those below it are thermally populated.
For electric-dipole coupling Ĥ′ = −μ̂·E(t). Every absorption, emission and scattering rate descends from this: squared coupling matrix element times density of final states.
Electronic: Δl = ±1, ΔS = 0. Vibrational: Δv = ±1 with ∂μ/∂Q ≠ 0 (IR) or ∂α/∂Q ≠ 0 (Raman); mutual exclusion in centrosymmetric molecules. Spin: ΔmI = ±1 (NMR), ΔmS = ±1 (EPR). The cos²θ law underlies LD, anisotropy and κ².
Spontaneous emission scales with ν³: radiative decay dominates in the UV–visible (ns) but is negligible for spin transitions, where relaxation takes over. f is dimensionless (ε in M⁻¹cm⁻¹, ν̃ in cm⁻¹); strong allowed bands have f ≈ 0.1–1.
Linear for dilute, non-scattering, non-interacting absorbers. Curvature flags aggregation, stray light, scattering or saturation. σ is the molecular absorption cross-section.
Dephasing and lifetime give Lorentzians (Δν = 1/2πτ: a 1 ns lifetime gives only 0.005 cm⁻¹). A static spread of environments gives Gaussians. Real bands are Voigt convolutions; 2D spectroscopy and hole burning separate the two.
Nuclei are frozen during an electronic transition. The Huang–Rhys factor S sets the vibronic progression; the Stokes shift ≈ 2Sħω is twice the reorganization energy (Marcus λ). Explains the absorption–emission mirror image.
Net polarization P at 298 K: ¹H at 600 MHz ≈ 5×10⁻⁵; electrons at 9.5 GHz ≈ 8×10⁻⁴; electronic transitions ≈ 1. This is why NMR needs mM samples, EPR µM, and fluorescence pM. Hyperpolarization (DNP) breaks the limit.
Absorption (χ″) and dispersion (χ′, refractive index) are two faces of one response. CD ↔ ORD, absorption ↔ refractive-index shifts in SPR, absorptive ↔ dispersive NMR phases.
T₁ restores populations; T₂ destroys coherence, with 1/T₂ = 1/2T₁ + 1/T₂′. The optical Bloch equations have the same form for two-level electronic systems, which is why pulse NMR, pulse EPR and nonlinear optics share one vocabulary.
Record a time-domain signal (NMR free induction decay, FTIR interferogram, pulse-EPR echo, 2D coherence times) and transform. Gains: multiplex (Fellgett) and throughput (Jacquinot). Resolution is set by acquisition length.
Line shapes come from dipole correlations, NMR relaxation from spectral densities J(ω), and DLS, FCS and anisotropy decays measure correlation functions directly. Fluctuation–dissipation links them all to equilibrium dynamics.
Time axis
Grey bars are biomolecular processes; coloured bars are the time windows techniques resolve directly. A technique can report on faster motions only as averages, and on slower ones only as static snapshots or by triggering.
Logarithmic time axis, 1 fs to about 3 h. Windows are typical, not hard limits. Select a coloured bar to open the technique.
Live calculators
Everything here runs in your browser. Defaults are realistic examples; change them to match your system.
E = 1/(1 + (r/R₀)⁶), with R₀ ∝ (κ²)1/6. Dashed curve: isotropic κ² = 2/3. The shaded band 0.5–1.5 R₀ is where E changes measurably.
Curves are area-normalized; the Voigt profile is their numerical convolution. Homogeneous dephasing time T₂ = 1/(πcΓL).
Rigid hydrated sphere: V = M(v̄ + δ)/NA with v̄ = 0.73 mL g⁻¹. D = kBT/6πηRh; θ = τc = ηV/kBT (Stokes–Einstein–Debye). FCS assumes w₀ = 0.25 µm; DLS assumes 633 nm, 90°, n = 1.33. Elongated or disordered proteins diffuse and tumble more slowly.
Question first
Start from the biological question, then pick a first-choice technique and at least one orthogonal check. Agreement between methods based on different physics is the strongest evidence you can offer a reviewer.
| Question | First choice | Orthogonal check | What you get |
|---|
Going deeper